LN2 DP LEVEL CONTROL & ALARM

❄️ DP-LLC Liquid Level Control & Alarm  30″ & 100″ models

Automated LN₂ Fill Control with Precision Differential Pressure Sensing

The DP-LLC series (models 30 & 100) are purpose-built for cryogenic LN₂ storage systems, offering reliable, automated fill solenoid control using differential pressure for level sensing.   Designed for lab, industrial, and biomedical environments, these controllers ensure safe, efficient liquid nitrogen management, featuring active alarms and intuitive user interfaces.

⚙️ Core Functions

  • Automated Fill Control
    Activates fill solenoid when LN₂ level drops below user-defined low setpoint
    Stops fill when high setpoint is reached.

  • Differential Pressure-Based Level Sensing
    Measures LN₂ level via pressure differential between tank and atmosphere
    Supports ranges up to 29″ (DP-LLC_30) or 100″ (DP-LLC_100)

  • Solenoid Valves: 24VDC Solenoid Valve (FILL)
  • Alarm Management

    • Low-Level Alarm: Triggers if level remains below setpoint for 7 minutes (user adjustable)

    • High-Level Alarm: Activates if level exceeds high setpoint for 2 minutes (user adjustable)

    • Source Alarm: Alerts if fill fails to reach high level within 2 hours (user adjustable)

  • Global Remote Alarm Contacts
    Dry contact outputs (NO/NC) for integration with external alert systems

  • Analog Output for Remote Monitoring
    0–5VDC signal proportional to LN₂ level.

🧠 Intelligent Features

  • User-Friendly Interface
    Backlit 16×2 LCD with tactile push-button controls
    Menu-driven setup for fill control, alarm thresholds, and calibration

  • Factory Calibration Access
    Secure menu for low/high level calibration, adjustable alarm timers, and unit measurement selection (in/cm)

  • Defog Timer
    Optional short-duration fill to clear vapor for visual inspection

  • Alarm Mute Timer
    Adjustable delay before audible alarms re-trigger (default: 10 min)

  • Level Filter
    Dampens sensor response to turbulence during fill

🔌 Electrical & Mechanical Specs

Parameter

Value

Power Input

100–240VAC, 50/60Hz, 1.5A

DC Adapter

24VDC @ 2A (Tri-Mag L6R48-240)

Power Consumption

9W (monitoring), 26W (fill active)

Sensor Range(s)

0–1.02 PSI (30″ range), 0-3.65 PSI (100″ range)

Accuracy

±1% FS, ±1 unit (inch/cm)

Display Resolution

0.5″ increments

Housing

ABS with acrylic panel, UL94V0 rated

Environmental Limits

Indoor use, 5–40°C, 80% RH max

CI_DP-LEVEL CONTROL MANUAL

 